Equipamentos multiusuário

FS5 Spectrofluorometer

 

 

 

Specification

Optics

All-reflective for a wavelength independent focus with high brightness (small focus) at the sample

Source

150 W CW Ozone-free xenon arc lamp

Monochromators

Czerny-Turner design with plane gratings for accurate focus at all wavelength and minimum stray light

Spectral Coverage - Excitation

230 nm - 1000 nm

Spectral Coverage - Emission

230 nm - 870 nm

Filter Wheels

Fully automated; included in both the excitation and emission monochromators

Bandpass - Excitation/Emission

0 to 30 nm, continuously adjustable

Wavelength Accuracy - Excitation/Emission

± 0.5 nm

Scan Speed - Excitation/Emission

100 nm/s

Integration Time

1 ms - 200 s

Emission Detector

Photomultiplier R928P, spectral coverage 200 nm - 870 nm, cooled and stabilised

Reference Detector

UV enhanced silicon photodiode

Transmission Detector

UV enhanced silicon photodiode

Water Raman Signal

>400,000 cps at 397 nm emission, excitation 350 nm, 5 nm bandpass, 1 s integration time

Signal-to-Noise Ratio of Water Raman Signal

SNR SQRT >6000:1

Dimensions

104 cm (w) x 59 cm (d) x 32 cm (h)

Weight

55 kg


FS5 Spectrofluorometer - Acesso para usuários externos

 


Pesquisadores vinculados a instituições de ensino/pesquisa podem cadastrar-se como usuários deste equipamento.

Para cadastrar um pesquisador responsável, preencher o arquivo "FS5 - Formulário de cadastro de pesquisador responsável".

Para enviar propostas de medidas, preencher o arquivo "FS5 - Formulário de proposta de medida".

Ambos devem ser enviados ao e-mail lna@if.usp.br para realização do cadastro formal e agendamento do equipamento.
 

Desenvolvido por IFUSP